Installation; Testing The System; Unpacking The Analyzer; Mounting The Analyzer - Teledyne 2000A-EU Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Thermal Conductivity Analyzer
Installation of the Model 2000A Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Installing the Sensor

6. Testing the system.

3.1
The analyzer is shipped ready to install and prepare for operation.
Carefully unpack the analyzer and inspect it for damage. Immediately
report any damage to the shipping agent.
The four gas fittings that mate with the 1/8 NPT gas ports on the
Model 2000A, are not included. They must be supplied by the customer.
3.2
The Model 2000A is for indoor use in a general purpose area. It is
NOT for hazardous environments of any type. It must be protected from:
• Direct sunlight
• Drafts of air
• Shock and vibration
• Temperatures below 30 °F (-1 °C) or above 110 °F (43 °C).
Locate the 2000A as close as possible, subject to the above conditions,
to the sample point to minimize effects of sample line lag time on the
analysis.
The standard model is designed for flush panel mounting. Figure 3-1 is
an illustration of the 2000A standard front panel and mounting bezel. There

Installation

Unpacking the Analyzer

Mounting the Analyzer

Teledyne Analytical Instruments
Installation 3
3-1