The Analyze Function - Teledyne 2000A-EU Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Thermal Conductivity Analyzer
Ranges that work well together are:
Range schemes that are to be avoided include:
Figure 4-2 illustrates these schemes graphically.
4.7
Normally, all of the functions automatically switch back to the Analyze
function when they have completed their assigned operations. Pressing the
Escape button in many cases also switches the analyzer back to the Analyze
function. Alternatively, you can press the Analyze button at any time to return
to analyzing your sample.
The Analyze function screen shows the impurity concentration and the
application gases in the first line, and the range in the second line. In the lower
Ranges that have the same lower limits but upper limits that differ
by approximately an order of magnitude
Ranges whose upper limits coincide with the lower limits of the
next higher range
Ranges where there is a gap between the upper limit of the range
and the lower limit of the next higher range.
Ranges that overlap
Ranges whose limits are entirely within the span of an adjoining
range.
Figure 4-2: Examples of Autoranging Schemes

The Analyze Function

Teledyne Analytical Instruments
Operation 4
4-21