Manual Mode Zeroing - Teledyne 2000A-EU Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Thermal Conductivity Analyzer
Then, and whenever Slope is less than 0.01 for at least 3 min, instead of
Slope you will see a countdown:
software steps in the zeroing process that the system must complete, AFTER
settling, before it can go back to Analyze . Software zero is indicated by
in the lower right corner.
The zeroing process will automatically conclude when the output is within
the acceptable range for a good zero. Then the analyzer automatically returns to
the Analyze mode.
Press Zero to enter the Zero function. The screen that appears allows you
to select between automatic or manual zero calibration. Use the
toggle between
blinking, on the display.
Press Enter to begin the zero calibration. After a few seconds the first of
three zeroing screens appears. The number in the upper left hand corner is the
first-stage zero offset. The microprocessor samples the output at a predeter-
mined rate.
The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–Zero
and F–Zero, use the
possible to zero. Then, press Enter .
S–Zero starts. During S–Zero, the Microcontroller takes control as in Auto
Mode Zeroing, above. It calculates the differences between successive sam-
plings and displays the rate of change as
second (ppm/s).
Note: It takes several seconds for the true Slope value to display.
Wait about 10 seconds. Then, wait until Slope is sufficiently
close to zero before pressing Enter to finish zeroing.
4.4.1.2

Manual Mode Zeroing

and
zero settling. Stop when
keys to adjust displayed
Teledyne Analytical Instruments
,
, and so fourth. These are
value as close as
a value in parts per million per
Operation 4
keys to
appears,
4-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents