The Range Function - Teledyne 2000A-EU Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

4 Operation
Value of the alarm setpoint, AL–1 ####
Out-of-range direction, HI or LO
Defeated? Dft:Y/N (Yes/No)
Failsafe? Fs:Y/N (Yes/No)
Latching? Ltch:Y/N (Yes/No).
To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the
change the number. Holding down the key speeds up the
incrementing or decrementing.
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the
keys to change the parameter.
Once the parameters for alarm 1 have been set, press Alarms
again, and repeat this procedure for alarm 2 (AL2).
To reset a latched alarm, go to
times or
Go to
(Toggle it to
4.6
The Range Select Function
The Range function allows you to manually select the concentration range
of analysis (
In the
(concentration) limits of each Range, and select a single, fixed range to run.
CAUTION: If this is a linearized application, the new range must
be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.
In the
screen, you are further allowed to select which gas applica-
tion (PREVIOUSLY defined in System function) to run.
4-18
two times. (Toggle it to and then back to )
–OR –
and then press either
and back to )
), or to select automatic range switching (
screen, you are further allowed to define the high and low
Teledyne Analytical Instruments
arrow keys to
arrow
and then press either
two times or
two times.
).
Model 2000A-EU