Siemens SOMATOM Manual page 373

Emotion 6/16-slice configuration
Hide thumbs Also See for SOMATOM:
Table of Contents

Advertisement

Vascular
Emotion 16
kV
Effective mAs/
Quality ref. mAs
Rotation time
Acquisition
Slice collimation 0.6 mm
Slice width
Feed/Rotation
Pitch Factor
Increment
Kernel
CTDI
Vol
Effective dose
Emotion 6
kV
Effective mAs/
Quality ref. mAs
Rotation time
Acquisition
Slice collimation 1.0 mm
Slice width
Feed/Rotation
Pitch Factor
Increment
Kernel
CTDI
Vol
Effective dose
CarotidAngio
110
70
0.6 sec.
16 x 0.6 mm
5.0 mm
14.4 mm
1.50
5.0 mm
B30s
5.60 mGy
Male: 1.28 mSv
Female: 1.41 mSv
CarotidAngio
110
70
0.8/0.6 sec.
6 x 1.0 mm
5.0 mm
9.0 mm
1.50
5.0 mm
B31s
5.81 mGy
Male: 0.57 mSv
Female: 0.62 mSv
nd
2
reconstr.
1.0 mm
0.7 mm
B20s
nd
2
reconstr.
1.25 mm
0.8 mm
B20s