Siemens SOMATOM Manual page 524

Emotion 6/16-slice configuration
Hide thumbs Also See for SOMATOM:
Table of Contents

Advertisement

Children
Contrast medium IV injection
Start delay
Flow rate
Total amount
Hints
• Children, who are older than age 6, should be
scanned with an adult protocol as the skull by this
time is fully grown.
• When bone structure is of interest, use kernel C60s
for image reconstruction.
• An advanced algorithm allows for improved head
image quality, without additional post-processing.
• In order to optimize image quality versus radiation
dose, scans are provided within a maximum scan
field of 300 mm with respect to the iso-center. No
recon job with a field of view exceeding those limits
will be possible. Therefore, patient positioning has to
be performed accurately to ensure a centered loca-
tion of the skull.
• To work without CARE Dose 4D use for children
< 6 month 90 mAs
6 month-3 years 150 mAs
3-6 years 220 mAs.
524
exam dependent
dependent upon needle size/
Access site
1 – 2 ml per kg of body weight

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents