Siemens SOMATOM Manual page 382

Emotion 6/16-slice configuration
Hide thumbs Also See for SOMATOM:
Table of Contents

Advertisement

Vascular
Contrast medium IV injection
Start delay
Flow rate
Total amount
Hints
• CARE Bolus may be used to optimize the bolus tim-
ing.
• Set the ROI for monitoring scan in the aortic arch
with triggering threshold of 120 HU, or use manual
triggering.
• Thick MPRs and ThinMIPs can be created very quickly
on the 3D Task Card by just clicking on the appropri-
ate icons. The thickness of these reconstructed
images can be defined by clicking on the icons with
the right mouse to open the entry field.
• Editing is necessary for SSD display of the thoracic
aorta without the bone.
382
10 – 25 sec.
2.5 ml/sec.
80 ml