Two combined ranges are predefined, head with neck
and thorax with abdomen.
For SOMATOM Emotion 16-slice configuration:
A scan range of 12/14 cm will be done in 13.36/9.68
sec.
Emotion 16
kV
Effective mAs/
Quality ref. mAs
Rotation time
Acquisition
Slice collimation
Slice width
Feed/Rotation
Pitch Factor
Increment
Kernel
CTDI
Vol
Effective dose
Male:
Female:
426
Head
130
240
1.0 sec.
16 x 1.2 mm
1.2 mm
6.0 mm
10.6 mm
0.55
6.0 mm
H31s
57.60 mGy
2.63 mSv
2.88 mSv
Neck
130
150
1.0 sec
16 x 1.2 mm
1.2 mm
5.0 mm
19.2 mm
1.00
5.0 mm
B50s
16.80 mGy
3.08 mSv
3.37 mSv