Body Kernels - Siemens SOMATOM Manual

Emotion 6/16-slice configuration
Hide thumbs Also See for SOMATOM:
Table of Contents

Advertisement

Thorax

Body Kernels

• As standard kernels for body tissue studies B30s or
B40s are recommended; softer images are obtained
with B20s or B10s (extremely soft). The kernels B31s
or B41s have about the same visual sharpness as
B30s, respectively, B40s, the image appearance,
however, is more agreeable due to a "fine-grained"
noise structure; quite often, the low contrast detect-
ability is improved by using B31s, B41s instead of
B30s, B40s.
• For higher sharpness, as is required e.g. in patient
protocols for cervical spine, shoulder, extremities,
thorax, lung, the kernels B50s, B60s, B70s, B80s are
available.
• A special High Resolution lung kernel is supplied
with B90s; by using kernel B90s, practically the same
image impression is obtained as with HCE-filtering of
a B40s image.
231