Dose Information
For spiral scan protocols, the indicated mAs is the
effective mAs per image. The correlation between tube
current and effective mAs of spiral scans on a Multi-
slice CT scanner is expressed by the following formula:
Effective mAs = mA x RotTime
Pitch Factor = Feed per Rotation
nrow x Slice collimation
mA = effective mAs
RotTime
where Slice collimation refers to the collimation of one
detector row, and nrow is the number of used detector
rows.
46
Pitch Factor
x Pitch Factor