Respiratory Gating
For proper belt positioning please follow the instruc-
tions below:
Step 1
patient
load cell fixing belt
load cell
The blue part should apply to the back of the
patient.
Step 3
Securely fasten the fixing belt around the
patient.
For further information on the respiratory gating sys-
tem AZ-733V (Anzai Medical, Japan), please refer to
the Operational manual "Respiratory Gating" or Anzai
User's manual.
Step 2
patient´s abdomen
the dia-
phragm
load cell
setting
point
navel
Wind the load cell fixing belt around the
body so that the pocket (load cell part)
applies ontojust below the diaphragm.
Step 4
load cell connection cable
Be careful load cell connection cable may
not entangle with the patient.