Vascular
Emotion 16
kV
Effective mAs/
Quality ref. mAs
Rotation time
Acquisition
Slice collimation 0.6 mm
Slice width
Feed/Rotation
Pitch Factor
Increment
Kernel
CTDI
Vol
Effective dose
Emotion 6
kV
Effective mAs/
Quality ref. mAs
Rotation time
Acquisition
Slice collimation 1.0 mm
Slice width
Feed/Rotation
Pitch Factor
Increment
Kernel
CTDI
Vol
Effective dose
HeadAngio
110
70
0.6 sec.
16 x 0.6 mm
3.0 mm
7.7 mm
0.80
3.0 mm
H31s
12.46 mGy
Male: 0.43 mSv
Female: 0.48 mSv
HeadAngio
110
70
0.8/0.6 sec.
6 x 1.0 mm
3.0 mm
5.1 mm
0.85
3.0 mm
H31s
12.88 mGy
Male: 0.37 mSv
Female: 0.42 mSv
nd
2
reconstr.
1.0 mm
0.7 mm
H20s
nd
2
reconstr.
1.25 mm
0.8 mm
H20s