Abdomen
AbdMultiPhase/
AbdMultiPhase06s
Indications:
Combination of three phase study including liver, pan-
creas and kidney.
For SOMATOM Emotion 16-slice configuration:
A range of 20 cm including liver, pancreas and kidney:
arterial phase acquired in 5.37 sec.
Emotion 16
kV
Effective mAs/
Quality ref. mAs
Rotation Time
Acquisition
Slice collimation
Slice width
Feed/Rotation
Pitch Factor
Increment
Kernel
CTDI
Vol
Effective dose
280
Non Contrast
110
75
0.6 sec.
16 x 1.2 mm
1.2 mm
5.0 mm
28.8 mm
1.50
5.0 mm
B41s
5.40 mGy
Male: 2.34 mSv
Female: 2.89 mSv