Adding Respiration Artifact To The Bp Signal; Numeric Codes For Dynamic-Wave Settings - Fluke Biomedical MPS450 Operator's Manual

Multiparameter simulator
Hide thumbs Also See for Biomedical MPS450:
Table of Contents

Advertisement

MPS450
Operators Manual
Alternatively, in the Numeric-Control Mode, press the number keys for a dynamic-wave
setting according to Table 2-25, and select RUN:
BP1 Arterial 120/80
BP1 Radial Art 120/80
BP1 Left Vent 120/00
BP1 Right Vent 25/00
BP2 Arterial 120/80
BP2 Radial Art 120/80
BP2 Left Vent 120/00
BP2 Right Vent 25/00
BP2 Pulmonary Art 25/10
BP2 Pulm Art Wedge 10/2
BP2 Left Atrium 14/4
BP2 Right Atrium CVP 15/10
BP3 Arterial 120/80
BP3 Radial Art 120/80
BP3 Left Vent 120/00
BP3 Right Atrium CVP 15/10
BP3 Pulmonary Art 25/10
BP3 Pulm Art Wedge 10/2
BP3 Left Atrium 14/4
BP3 Right Vent 25/00

Adding Respiration Artifact to the BP Signal

The MPS450 adds respiration artifact to any BP-channel dynamic wave, including the
Swan-Ganz-wave series. The artifact is turned on separately for each of the four BP
channels.
The following peak-to-peak changes (mmHg) are caused by the respiration wave, with
values independent of the programmed respiration amplitude:
for BP channel 1, a delta change from 0 to 16;
2-32
Table 2-25. Numeric Codes for Dynamic-Wave Settings
Dynamic-Wave Setting
NumericCode
060
061
062
063
070
071
072
073
074
075
076
077
080
081
082
083
084
085
086
087
gje031.eps