MPS450
Operators Manual
Apnea Simulation
Table 4-24 lists the numeric control codes for the apnea simulation functions.
Control Code
152
153
154
150
151
Blood Pressure Functions
The Blood Pressure Functions consists of Blood-Pressure Sensitivity, Blood Pressure
Zeroing, BP Channel 1: Static-Pressure Levels, BP Channel 2: Static-Pressure Levels, BP
Channel 3: Static-Pressure Levels, BP Channel 4: Static-Pressure Levels, BP Channel 1:
Dynamic Waveforms, BP Channel 2: Dynamic Waveforms, BP Channel 3: Dynamic
Waveforms, BP Channel 4: Dynamic Waveforms, BP Channel 1: Respiration Artifact,
BP Channel 2: Respiration Artifact, BP Channel 3: Respiration Artifact, and BP
Channel 4: Respiration Artifact.
Blood-Pressure Sensitivity
Table 4-25 lists the numeric control codes for the blood-pressure sensitivity functions.
Control Code
069
068
132
133
(N/A)
Blood Pressure Zeroing
Table 4-26 lists the numeric control codes for the blood-pressure zeroing functions.
Control Code
007
BP Channel 1: Static-Pressure Levels
Table 4-27 lists the numeric control codes for the BP channel 1: static-pressure levels
functions.
4-16
Table 4-24. Numeric Codes for Apnea Simulation Actions
Numeric
Start apnea for 12 seconds.
Start apnea for 22 seconds.
Start apnea for 32 seconds.
Turn on apnea continuously.
Turn off apnea.
Table 4-25. Numeric Codes for Blood-Pressure Sensitivity Actions
Numeric
Set blood-pressure sensitivity to 5 V/V/mmHg.
Set Set blood-pressure sensitivity to 40
V/V/mmHg.blood
Store blood-pressure sensitivity at 5 V/V/mmHg.
Store blood-pressure sensitivity to 40 V/V/mmHg.d
Store the blood-pressure sensitivity
Table 4-26. Numeric Codes for Blood-Pressure Zeroing Actions
Numeric
Zero all four blood pressure channels.
Action
Action
Action
Remote Control
Entry
A12
A22
A32
AON
AOFF
Remote Control
Entry
BPSNS5
BPSNS40
(N/A)
(N/A)
BPSNSSTORE
Remote Control
Entry
ZALL