Alternatively, in the Numeric-Control Mode, press the number keys for an ECG-
amplitude setting according to Table 2-2, and select RUN:
Table 2-2. Numeric Codes for ECG Amplitude Settings
Amplitude setting
0.05 mV
0.10 mV
0.15 mV
0.20 mV
0.25 mV
0.30 mV
0.35 mV
0.40 mV
0.45 mV
0.50 mV
1.00 mV
1.50 mV
2.00 mV
2.50 mV
3.00 mV
3.50 mV
4.00 mV
4.50 mV
5.00 mV
5.50 mV

Adult and Pediatric ECG

The MPS450 simulates an adult ECG waveform with QRS duration of 80 ms, or a
pediatric waveform with QRS duration of 40 ms.
Action
1. Press the top-menu key labeled NSR. The LCD screen identifies the current patient
type either as adult (ADULT) or as pediatric (PEDS).
2. Select > to move to the LCD screen PT TYPE.
3. Select PT TYPE to toggle to the desired patient type.
4. The patient-type setting identified on the LCD screen is active when displayed and
remains active until the setting is changed.
Alternatively, in the Numeric-Control Mode, press the number keys for a patient-type
setting according to Table 2-3, and select RUN:
Numeric code
252
253
254
255
256
257
258
259
260
261
262
263
264
265
266
267
268
269
270
271
2
Cardiac Functions
ECG Functions
2-5