Numeric Codes For Iup Wave Settings; Numeric Codes For Iup Contraction Period Settings - Fluke Biomedical MPS450 Operator's Manual

Multiparameter simulator
Hide thumbs Also See for Biomedical MPS450:
Table of Contents

Advertisement

UNIF ACC (uniform acceleration).
5. Select RUN. On the LCD screen, RUNNING flashes while the simulation is active.
6. A manually generated IUP ends after one contraction. To repeat the simulation, select
RUN again.
7. Periodically generated IUPs repeat at the selected period. To terminate a periodic-
IUP simulation, select END.
Alternatively, in the Numeric-Control Mode, press the number keys for an IUP-wave
simulation using one of the codes in Table 2-33, and select RUN. (The selected IUP is
generated according to the specific period that was last selected, whether manual, or once
every 2, 3, or 5 minutes.)
Table 2-33. Numeric Codes for IUP Wave Settings
IUP Wave Setting
IUP OFF
IUP UDEC (uniform deceleration)
IUP EDEC (early deceleration)
IUP LDEC (late deceleration)
IUP UACC (uniform acceleration)
To select a contraction period in the Numeric-Control Mode, press the number keys for
one of the settings found in Table 2-34, and select RUN. (The selected period is applied
to a run of the specific IUP-wave simulation that was last selected, whether uniform
deceleration, early deceleration, late deceleration, or uniform acceleration.)
Table 2-34. Numeric Codes for IUP Contraction Period Settings
IUP Contraction Period Setting
IUP MANUAL
IUP 2 MIN
IUP 3 MIN
IUP 5 MIN
Numeric Code
400
401
402
403
404
Numeric Code
405
406
407
408
Cardiac Functions
Fetal / Maternal ECG (Option)
2
2-41