Connection The Mps450; Ecg Jack Labeling - Fluke Biomedical MPS450 Operator's Manual

Multiparameter simulator
Hide thumbs Also See for Biomedical MPS450:
Table of Contents

Advertisement

MPS450
Operators Manual

Connection the MPS450

The right side of the MPS450 features connections for linking to blood-pressure, cardiac-
output, and temperature monitors. In addition, there is an auxiliary connection for future
expansion. The bottom of the MPS450 features an RS-232 serial port and a connection
for a battery eliminator. Prewired cables compatible with all major manufacturers'
monitors are available for simulating functions related to blood pressure, temperature,
and cardiac-output injectate. (Call your sales representative for a complete list.)
The left side of the MPS450 features a full set of universal ECG jacks, enabling the
connection of any 3-, 5-, or 12-lead ECG device. AHA and IEC color-coded dots run
along the left side of the face of the unit as an aid in connecting the corresponding U.S.
and international patient leads to the proper universal ECG jacks on the MPS450:
RA or R
LA or L
RL or N
LL or F
V1, V2, V3, V4, V5, and V6
C1, C2, C3, C4, C5, and C6
1-8
Figure 1-1. MPS450 Patient Simulator
Table 1-2. ECG Jack Labeling
Label
Meaning
Right arm
Left arm
Right leg (reference or ground)
Left led
V Leads (U.S. and Canada). Also referred to as
pericardial, precordial, or unipolar chest leads.
Chest leads (International)
gje041.eps

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents