To avoid severe electrical shock, disconnect the power
source before replacing fuses.
1. Make sure that the power switch is in the OFF position (O) and disconnect
the power cord. The Analyzer is now safely powered down.
2. Pry the left side of the fuse cover up from the power inlet assembly, using
a small flat-blade screwdriver. The fuse cover latches on the right side and
remains attached to the Analyzer.
3. Remove a fuse carrier with your fingers, lifting it from the power inlet
housing.
To avoid damage to the Analyzer, do not use any sharp instrument to
lift a fuse carrier.
4. Replace the fuse in the carrier with one of the same amperage and voltage
ratings.
Installation, Setup, and Maintenance
Figure 2-1. Fuse Cover
W X Warning
Note
2
Maintenance
exz16.eps
2-11