Fluke RF303 Operator's Manual page 50

Electrosurgical analyzer
Table of Contents

Advertisement

RF303
Operators Manual
To measure the ESU's signal HF leakage:
1. If the mA indicator lamp on the panel is lit, go to Step 2. If it is not lit,
press the MODE SELECT (ENT) button to select mA and go to Step 2.
2. Activate the ESU and view the signal HF leakage measurement in
milliamperes on the display. Here is an example.
To end HF leakage test number two:
Deactivate the ESU to end the HF leakage test.
To avoid damage to the Analyzer and the ESU under test,
do not exceed the Analyzer's duty cycle, which is 30
seconds ON during any continuous one-minute period.
3-14
ENT
MODE SELECT
ENT
MODE SELECT
W Caution
WATTS
mA
OHMS
OHMS SELECT
WATTS
mA
OHMS
OHMS SELECT
exz07.eps
exz18.eps