Fluke RF303 Operator's Manual page 62

Electrosurgical analyzer
Table of Contents

Advertisement

RF303
Operators Manual
Table 3-2. Available Duplex Mode Commands (cont.)
Command
3-26
Description
Prompts the Analyzer to
return all measurements
of the applied signal of
the ESU.
{MODE=OUTPUT}
Prompts the Analyzer to
measure the applied
signal current of the
ESU.
{MODE=RFLKG}
Prompts the Analyzer to
return the current test
load.
{MODE=RARF}
Parameters
Return Value
Required
None
I=AAAA(mA),
CF=XX.X,
KVPP=XXX,
PWR=PPP.P(w)
Where AAAA is the
Current reading in
milliamps, XX.X is the
Crest Factor, XXX is
the peak-to-peak
voltage in kVolts and
PPP.P is the power in
watts
None
I=AAAA(mA)
Where AAAA is the
Current reading in
milliamps.
None
REM RESISTANCE=
RRR
Where RRR is the
load value in ohms.
(Response)