To measure the ESU's power output:
1. Activate the ESU.
2. If the WATTS indicator lamp on the panel is lit, go to Step 3. If it is not
lit, press the MODE SELECT (ENT) button to select WATTS and go to
Step 3.
3. View the power measurement on the display. Here is an example:
To measure the ESU's current output:
1. While the ESU is still activated, press the MODE SELECT (ENT) button
until the milliamperes (mA) indicator lamp lights.
2. View the ESU's current output (in mA). Below is an example.
To end the generator output test:
Deactivate the ESU.
To avoid damage to the Analyzer and the ESU under test,
do not exceed the Analyzer's duty cycle, which is 30
seconds ON during any continuous one-minute period.
3-6