RF303
Operators Manual
All responses are uppercase, followed by CR and LF.
Measurement readings from the Analyzer are returned identically as
they would appear on the Analyzer's display; that is, as numbers, with
or without a decimal point, and followed by any appropriate unit of
measurement.
Errors always return a response beginning with: ?ERR=XX (plus the
description). Table 3-1 lists each of the error responses.
Error Response
?ERR=01,BAD OR MISSING
PARAMETERS
?ERR=02,ILLEGAL NUMBER OF
PARAMETERS
?ERR=03,PARAMETER OUT OF
RANGE OR SYNTAX ERROR
?ERR=04,COMMAND NOT
AVAILABLE FROM THIS MODE
?ERR=05, DEACTIVATE ESU
BEFORE ADJUSTING
?ERR=06,ACCESS DENIED
Table 3-2 describes various commands available in Duplex Mode.
3-22
Table 3-1. Error Responses
The received command requires
parameters, and there were none, or
they were not recognized.
The command did not include the
correct number of parameters.
One or more of the parameters were
out of the correct range, or there was
an error in the syntax of the
command.
The Model Analyzer is not in the
special sub mode within the remote
that is required for the command.
There is an ESU signal present at the
meter inputs that is not allowed for
the command to be executed.
The command is a calibration
command and the calibration jumper
is not connected
Explanation