44.1/48 Base Sample Rate Menu - Pacific Microsonics HDCD Model Two User Manual

A/d converter d/a converter digital processor
Table of Contents

Advertisement

Reference
45
®
Model Two HDCD
User's Manual
Page two of the Setup Menu:
Button #1 (SYSTEM) selects the System Menu, button #2 (MONITOR) selects the Monitor
Setup Menu, button #3 (METER) selects the Meter Setup Menu, button #4 (STORE_P) se-
lects the Store Presets Menu, and button #5 (MORE...) selects page three of the Setup
Menu. Descriptions of the above menus begin on pages 59, 61 and 63.
Page three of the Setup Menu:
Button #1 (DOWNLD) activates the system software download procedure. Prior to initiat-
ing download, the following message is displayed, "Download new system file? Press EN-
TER to confirm, or any other key to cancel". Pressing ENTER brings up the Select Baud Rate
Menu to start the download process. Refer to System Software Download on page 65.
Button #2 (TEST) selects the Test Menu, button #3 (VERSION) selects the Software Version
Menu, button #5 (MORE...) returns to page one of the Setup Menu.
Descriptions of the above menus begin on page 64.

44.1/48 Base Sample Rate Menu

The 44.1/48 Base Sample Rate Menu is reached from the Operating Menu by selecting
(SETUP/MODE).
Button #1 (44.1_X) selects 44.1 kHz as the base sample rate for all Model Two operating
P A C I F I C
M I C R O S O N I C S
I N C .
Rev. C - January 2002
®