To perform the calibrations:
1 Turn the control wheel to display the Calibration menu.
***MAIN MENU****
Calibration
2 Click the control wheel to enter the Calibration menu.
3 Turn the control wheel to select the calibration you wish to
perform.
4 For the automatic calibrations: Click the control wheel to perform
the calibration. During the calibration, the progress bar in the on-
screen display is filled out.
After successful calibration, the on-screen display shows the word
"Successfully".
After unsuccessful calibration, the on-screen display shows the
word "Failed". Please refer to chapter 5 : Troubleshooting for
more information.
5 For manual calibrations:
a) Click the control wheel to enter the menu.
b) Turn the control wheel to change the contrast or brightness
c) Click the control wheel to confirm the change of the value.
6 After calibration, click and hold the control wheel for a few
seconds to exit the Calibration menu.
The different calibration functions:
Total Autoset ............This function performs the automatic detection
Phase & level ...........Automatic calibration of the video sampling
Auto level..................Automatic calibration of the video levels (black
Auto phase ...............Automatic calibration of the video sampling
Manual phase ...........Manual adjustment of the video sampling phase
Manual black ............Manual adjustment of the video black level
Manual white ............Manual adjustment of the video white level
User manual MIVD 1218
value.
of the number of pixels (horizontal and vertical)
and all other calibration functions (described
below) one after the other. This function takes a
few minutes.
phase and the video levels (black and white)
and white)
phase
(brightness cal setting)
(contrast cal setting)