Calibration process
N O T E
Agilent U1251A/U1252A User's and Service Guide
Now you can use 1234 as the security code. If you want to
enter a new security code, see
calibration security code from the front panel"
Make sure you record the new security code.
To complete a full instrument calibration use following
general procedures:
1 Read
"Test Considerations"
2 Perform the verification tests to characterize the
instrument (incoming data).
3 Unsecure the instrument for calibration (see
Security"
on page 119).
4 Perform the adjustment procedures (see
Consideration"
on page 133).
5 Secure the instrument against calibration.
6 Note the new security code and calibration count in the
instrument's maintenance records.
Make sure to quit the Adjustment Mode before turning off the instrument.
Performance Tests and Calibration
"To change the instrument
on page 118.
"Adjustments
7
on page 129.
"Calibration
131