Frequency Counter Specifications for U1252A
Table 7-15 Frequency counter (divide by 1) specifications
Range
99.999 Hz
999.99 Hz
9.9999 kHz
99.999 kHz
999.99 kHz
9.9999 MHz
Table 7-16 Frequency counter (divide by 100
Range
9.9999 MHz
99.99 MHz
Notes for frequency counter specifications:
1 All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from
picking up external noise is critical for minimizing measurement errors. For non-square wave signals, an additional 5
counts need to be added.
2 The maximum measurement level is < 30 Vpp.
3 The minimum measurement frequency of low frequency is set by power-on option to speed up the measurement rate.
4 Shown on the secondary display
160
Accuracy
± (% of reading + No.
Resolution
of Least Significant
0.001 Hz
0.02% + 3
0.01 Hz
0.0001 kHz
0.001 kHz
0.002% + 5
0.01 kHz
0.0001 MHz
[4]
) specifications
Accuracy
± (% of reading + No.
Resolution
of Least Significant
0.0001 MHz
0.002 % + 5,
0.001 MHz
Sensitivity
Digit)
[1]
100 mV R.M.S.
200 mV R.M.S.
< 985 kHz
Sensitivity
Digit)
400 mV R.M.S.
600 mV R.M.S.
< 20 MHz
Agilent U1251A/U1252A User's and Service Guide
Min. Input Frequency
0.5 Hz
Min. Input Frequency
1 MHz