Valid Adjustment Input Values; Table 6-3 Valid Adjustment Input Values - Agilent Technologies U1251A User's Manual And Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1251A:
Table of Contents

Advertisement

7
Performance Tests and Calibration

Valid adjustment input values

Table 6-3 Valid adjustment input values

Function
V
(for U1251A)
(for U1252A)
Ω
/ TEMP
134
Adjustment can be accomplished using the following input
values below.
Range
5 V, 50 V, 500 V, 1000 V
5 V, 50 V, 500 V, 1000 V
5 V, 50 V ,500 V, 1000 V
50 mV, 500 mV, 1000 mV
500
A, 5000
µ
µ
50 mA, 440 mA, 5 A, 10 A
500 Ω, 5 kΩ, 50 kΩ,500 kΩ, 5 MΩ,
50 MΩ
10 nF, 100 nF, 1000 nF, 10
1000
F, 10 mF / 0 °C
µ
A
F, 100
F,
Make sure to provide 0 °C with ambient
µ
µ
Agilent U1251A/U1252A User's and Service Guide
Valid Amplitude Input Values
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
compensation