7
Performance Tests and Calibration
Step
Test Function
7
Turn the rotary switch to the
TEMP/
position
8
Press
button to go to
TEMP mode
Note: Ensure the multimeter is turned on and stabilized for at least 60 minutes, with the K-type thermocouple
connected between the multimeter and the calibrator output terminal
9
Turn the rotary swich to the
mA position
140
Cal Range
Input
Open
Input terminal
open (Remove any
test leads and
Short Plugs from
the input terminal)
10 nF
3 nF
10 nF
100 nF
10 nF
100 nF
1000 nF
100 nF
1000 nF
10 μF
10 μF
100 μF
100 μF
1000 μF
1000 μF
10 mF
10 mF
N/A
0 ºC
OPEN
Input terminal
open (Remove any
test leads and
Short Plugs from
the input terminal)
500 μA
300 μA
5000 μA
3000 μA
Agilent U1251A/U1252A User's and Service Guide
Cal Item
U1251A
U1252A
oPEn
oPEn
03.000 nF
03.000 nF
10.000 nF
10.000 nF
010.00 nF
010.00 nF
100.00 nF
100.00 nF
0100.0 nF
0100.0 nF
1000.0 nF
1000.0 nF
10.000 μF
10.000 μF
100.00 μF
100.00 μF
1000.0 μF
1000.0 μF
10.000 mF
10.000 mF
0000.0 ºC
0000.0 ºC
oPEn
oPEn
300.00 μA
300.00 μA
3000.0 μA
3000.0 μA