Optional AC Current Performance Verification Test

34405A User's and Service Guide
Configuration: AC Current (CONFigure:CURRent:AC)
1 Select the AC Current function.
2 Select each range in the order shown below. Provide the
indicated input voltage and frequency. Compare
measurement results to the appropriate test limits shown
in the table. (Be certain to allow for appropriate source
settling.)
Table 12
Optional AC Current Performance Verification Test
Current
Input Frequency Function
10mA
20Hz
10mA
45Hz
10mA
1kHz
10mA
10kHz
1A
1kHz
100mA
1kHz
10mA
1kHz
[1] Select 5½- digit measurement resolution
Performance Tests and Calibration
[1]
Range
AC Current
10mA
10mA
10mA
10mA
1A
1A
1A
4
Error from Nominal
1 year
± 0.16mA
± 0.16mA
± 60µA
± 0.22mA
± 6mA
± 1.5mA
± 1.05mA