Cleaning the Containers

Cleaning the Containers
Procedures for emptying, cleaning and refilling the Wash Container and the
Waste Container are described below.
To empty and clean the waste container:
1. Unscrew the cap of the Waste Container. Lay the cap and level
sensor assembly on a paper towel.
2. Discard the waste in accordance with local regulations.
3. Rinse the waste container with deionized water. Discard the rinse
water.
4. Place the level sensor and tube back into the Waste Container
and screw the cap on.
To refill the wash container:
1. Separate the dispense line quick-connector in the cap of the
Wash Container. Unscrew the cap and lay it and the Pick-Up
Filter on a clean paper towel.
2. Fill the Wash Container with the wash fluid that is being used.
3. Place the Pick-Up Filter into the Wash Container and screw the
cap on. Reconnect the dispense line quick-connector.
70
Note: If desired, the waste container can be
disinfected with a 10% (v/v) solution of
household bleach in water.
Note: If the wash solution is being changed,
discard the contents of the Wash Container
and thoroughly clean it before filling it with the
new wash solution.