Routine Maintenance Procedures
Replace the dispense (Wash) tubing.
Replace the aspirate (Waste) tubing.
Replace the in-line hydrophobic filter.
If the system will not be used for a period of time, the system should
be flushed with de-ionized water so that buffers and/or reagents
used to wash the plates are not deposited on the Wash Head or
other components of the system.
As required maintenance:
Pump Calibration (page 74)
68
Note: The dispense tubing and aspiration
tubing may need to be replaced more
frequently than every six months, depending
upon the frequency of use and the severity of
operating conditions.