To specify a Move cycle:
1. Display the header information for the selected wash protocol, as
outlined on page 41. The display will read:
Select option
2. Select MOVE. The display will read:
Start strip?
3. Using the keypad, enter the strip number where the move cycle
should start.
4. Press ENTER. The display will read:
End strip?
5. Using the keypad, enter the strip number where the move cycle
should end.
6.
Press ENTER. The display will read:
Cycles?
7. Using the keypad, enter the number of the specified move cycles
that are to be sequentially performed.
8. Press ENTER. The display will read:
Select option
ASPIRATE DISPENSE
9. Specify Aspirate, Dispense, Soak and Fill operations in the
sequence as they are to be carried out, as described in the
following sections.
10. When all of the operations for the Move cycle have been
sequentially specified, specify End Loop to designate the end of
the Move cycle.
Opsys MW™ User's Guide
Creating a New Wash Protocol
MOVE
PURGE
1
12
1
Note: Between 1 and 9 Move cycles can be
specified.