7. Enter the dispense height using the keypad and press ENTER.
The display will read:
8. Enter the desired value using the keypad and press ENTER. The
display will read:
9. Enter the desired value using the keypad and press ENTER. The
display will read:
10. Select YES to specify a sweep. The display will read:
32
Well top?
160
SHOW
Note: The default value is displayed. Select
SHOW to move the wash head to the default
well top position so that the height can be
checked. Then, determine the correct value as
outlined on page 30. Repeat the procedure as
needed.
ASP. HEIGHT?
380
SHOW
Note: The default value is displayed. Select
SHOW to move the wash head to the default
aspiration position so that the height can be
checked. Then, determine the correct value as
outlined on page 30. Repeat the procedure as
needed.
Sweep?
NO
YES
NO
All cycles?
NO
YES
NO
Note: If a sweep is specified, the wash pins
are moved up and down and across the
bottom of the well during aspiration to better
remove all droplets of liquid. If NO is selected,
the Wash height? prompt (Step 14) will be
displayed.
Note: Use the sweep function with Flat or "C"
bottom plates only. Sweeping aspirates fluid
more completely from the wells than a regular
aspiration. Sweep all cycles to get the most
thorough well aspiration when using multiple
aspiration cycles, but remember that each
sweep adds seconds to the protocol.