Page 2
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 3
(UPS) as related to Narkomed model numbers 4, 3, 2C, 2B, and GS. The APC Power Supply Kit numbers will vary depending on the model number of the Narkomed Machine. Kit number 4117870-001 is to be used on the Narkomed 4, 3, and 2C. Kit number 4117870-001 consists of the following parts:...
Page 4
Ensure that A/C outlets are facing the left side of the machine. See Figure 5. NOTE: Narkomed 4 model machines will require that cables be routed through corner notch between power supply assembly base plate and machine frame rails.
Page 5
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Replacement Procedure • Ensure circuit breakers are ‘in’. • Ensure AC power is applied to the machine. • Ensure system power switch is in the ‘on’ position. •...
Page 6
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Replacement Procedure KEYBOARD MOUSE POWER SUPPLY ASSEMBLY POWER COMMUNICATIONS POWER TOUCH (WALL) CONTROL MONITOR VIDEO DISPLAY FLAT PANEL POWER PROCESSOR UNIT (WALL) NETWORK SU00060 FIGURE 2. Typical 400 Mhz Processor Cable Connections...
Page 7
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Replacement Procedure KEYBOARD MOUSE POWER SUPPLY ASSEMBLY POWER COMMUNICATIONS POWER TOUCH (WALL) CONTROL MONITOR VIDEO DISPLAY FLAT PANEL POWER PROCESSOR UNIT (WALL) NETWORK SU00061 FIGURE 3. Typical 700 MHZ Processor Cable Connections...
Page 8
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Replacement Procedure SATURN POWER SUPPLY (4116687-001 &002) EXISTING HARDWARE NARKOMED POWER SUPPLY FRAME SPACER MACHINE POWER SU00034 SUPPLY FIGURE 4. Existing Saturn Power Supply Mounting Configurations on Narkomed 4, 3, and 2C...
Page 9
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Replacement Procedure MACHINE POWER SUPPLY NEW SATURN POWER SUPPLY ASSEMBLY LOCATION OF NOTCH CUTOUT FOR NARKOMED 4 CABLE ROUTING FIGURE 5. Replacement Saturn Power Supply Mounting Configuration on Narkomed 4, 3, and 2C...
Page 10
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Replacement Procedure KIT NUMBER 4117870-002 - NARKOMED 2B AND GS 1. Turn the system power switch to STANDBY. 2. Disconnect AC power from the machine and pull all circuit breakers to their ‘out’ position.
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Replacement Procedure • Ensure circuit breakers are ‘in’. • Ensure AC power is applied to the machine. • Ensure system power switch is in the ‘on’ position. •...
Page 12
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Replacement Procedure NEW SATURN POWER SUPPLY NEW SATURN POWER SUPPLY MOUNTING HARDWARE (TYPICAL FOUR PLACES) FIGURE 7. Replacement Saturn Power Supply Mounting Configuration on Narkomed 2B and GS...
Page 13
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 14
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS DrägerService is a division of Draeger Medical, Inc. 3122 Commerce Drive Telford, PA 18969 Tel: (215) 721-5402 (800) 543-5047 Fax: (215) 721-5784 Web: www.draegermedical.com Printed in the U.S.A.