PTR308 Automatic Recoater with Linear Proof Tester
Advanced Parameters
Advanced parameters are accessed from the toolbar since they are rarely modified. The PTR308 advanced
parameters are defined as follows:
Shrink Factor:
Shrink Time:
Lamp Set 1 Start:
Lamp Set 1 Stop:
Lamp Set 2 Start:
Lamp Set 2 Stop:
Lamp Set 3 Start:
Lamp Set 3 Stop:
Lamp Set 4 Start:
Lamp Set 4 Stop:
Lamp Set 5 Start:
Lamp Set 5 Stop:
Lamp Set 6 Start:
Lamp Set 6 Stop:
Lamp Set 7 Start:
Lamp Set 7 Stop:
Lamp Set 8 Start:
Lamp Set 8 Stop:
Page 20
The Shrink Factor is the amount of material injected into the recoat mold during the
early stages of curing. This is entered as a percentage of the initial Inject volume, and
is intended to compensate for material shrinkage during cure. Typical values for Shrink
Factor are 3% to 6%.
Shrink Time specifies the time period over which any additional material is injected into
the mold during cure. It is expressed as a percentage (%) of Cure Time.
Specifies the switch-on time of Lamp Set 1 as a percentage (%) of Cure Time.
Specifies the switch-off time of Lamp Set 1 as a percentage (%) of Cure Time.
Specifies the switch-on time of Lamp Set 2 as a percentage (%) of Cure Time.
Specifies the switch-off time of Lamp Set 2 as a percentage (%) of Cure Time.
Specifies the switch-on time of Lamp Set 3 as a percentage (%) of Cure Time.
Specifies the switch-off time of Lamp Set 3 as a percentage (%) of Cure Time.
Specifies the switch-on time of Lamp Set 4 as a percentage (%) of Cure Time.
Specifies the switch-off time of Lamp Set 4 as a percentage (%) of Cure Time.
Specifies the switch-on time of Lamp Set 5 as a percentage (%) of Cure Time.
Specifies the switch-off time of Lamp Set 5 as a percentage (%) of Cure Time.
Specifies the switch-on time of Lamp Set 6 as a percentage (%) of Cure Time.
Specifies the switch-off time of Lamp Set 6 as a percentage (%) of Cure Time.
Specifies the switch-on time of Lamp Set 7 as a percentage (%) of Cure Time.
Specifies the switch-off time of Lamp Set 7 as a percentage (%) of Cure Time.
Specifies the switch-on time of Lamp Set 8 as a percentage (%) of Cure Time.
Specifies the switch-off time of Lamp Set 8 as a percentage (%) of Cure Time.