Section 5
Certification Procedures for the Axiom
This section will outline test procedures to validate the performance of the Axiom.
Note: These are the only Labconco-approved test procedures for validating
the performance of the Axiom Biosafety Cabinet. Other test procedures may
yield different or inaccurate test results.
NOTE: In order to properly certify the Axiom, you must obtain an Axiom
Certifier Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by
contacting Labconco Customer Service Department at 800-821-5525.
Note: The ECM motor is programmed by Labconco to deliver a constant
volume of air throughout the life of the HEPA filters. Any changes to the
blower speed control should be small (less than 5 units).
Accessing the Certifier's Menu
Note: You will need to access this menu to do any of the following:
•
Adjust the blower speed
•
To reset the filter gauge
•
To calibrate the optional airflow sensor
Adjusting the Downflow, Inflow and Filter Life Gauges
Menu
1. Press the
2. Press the
▼
button until the
turn yellow).
OK/Mute
3. Press
4. Press the
▼
button until the
will turn white).
OK/Mute
5. Press
warning screen
6. You will see a
alter the BSC's settings.
OK/Mute
7. Press
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Calibration
to start the calibration procedure.
alerting you that you are about to
to input the password.
Product Service 1-800-522-7658
option is highlighted (it
S ervice
C urrent S ettings
R estore D efault
D iagnostics
C alibration
C onfiguration
or
t o choose,
O K t o select , ME N U t o ret urn
C alibrat ion
ATTE N TIO N !
The following screens are used
to alter the operation of the
L ogic . T hey should only be used
by a qualified certifier as part
of the certification process.
O K t o proceed, ME N U to return