Labconco Purifier Axiom 30441 Technical Manual And Specifications page 10

Type c1 biosafety cabinet
Hide thumbs Also See for Purifier Axiom 30441:

Advertisement

Quickstart
How do I recalibrate the
✓ blower speed
✓ filter life gauge
✓ airflow sensor reading
1. Press the
Menu
2. Press the
button until the
turn yellow).
OK/Mute
3. Press
4. Press the
button until the
will turn white).
OK/Mute
5. Press
warning screen
6. You will see a
alter the BSC's settings.
OK/Mute
7. Press
8. When requested for the password press
Timer
Timer
,
9. If the password is properly entered, the first certification screen
will display the supply blower motor's PWM setting. Press the
or
button to increase or decrease the blower speed as
needed. When the speed is set properly, press the
button to lock the new blower speed setting, and continue to the
filter life screen.
10
of the Axiom?
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Calibration
to start the calibration procedure.
alerting you that you are about to
to input the password.
OK/Mute
then
.
Product Service 1-800-522-7658
option is highlighted (it
Light
UV light
,
,
OK/Mute
C alibrat ion
ATTE N TIO N !
The following screens are used
to alter the operation of the
L ogic . T hey should only be used
by a qualified certifier as part
of the certification process.
O K t o proceed, ME N U to return
C alibrat ion
PAS S W O R D :