Technical Data
Influencing
Sphere of Influence
Quantity
Interference quantity max. 250 V
Common mode
interference
Interference quantity max. 250 V
voltage
50 Hz, 60 Hz sine
Interference quantity V ,
respective nominal value
Series-mode
of the measuring range,
interference
max. 250 V
, 50 Hz, 60 Hz, sine
voltage
Interference quantity max. 250 V
Real-Time Clock
Resolution
0.1 s
1 minute per month
Accuracy
Temp. Influence
50 ppm/K
Reference Conditions
Ambient temperature +23C 2 K
40 60%
Relative humidity
Meas. quantity
45 65 Hz
frequency
Meas. quantity
waveform
Sinusoidal, deviation between RMS and
rectified value < 0.1%
3.0 V 0.1 V
Battery Voltage
72
Response Time (multimeter functions)
Response Time (after manual range selection)
Measuring
Attenuation
dB
Range
Measured Quantity /
V
> 90 dB
Measuring Range
300 mV
> 80 dB
30 V
300 ... 3 M
300 V
> 70 dB
V
> 60 dB
V
> 60 dB
3 nF ... 30 F
Display
LCD panel (65 x 35 mm) with display of up to 3 measured values,
unit of measure, type of current and various special functions.
Display / char. height 7-segment characters
Number of places
Overflow display
Polarity display
LCD test
Digital Display
Response Time
V DC, V AC
1.5 s
A DC, A AC
2 s
30 M
5 s
Continuity
< 50 ms
1.5 s
C Pt100
Max. 3 s
Max. 2 s
>10 Hz
Max. 1.5 s
Main display:
Auxiliary displays: 7 mm
4¾-places,
"OL" or "–OL" appears
"–" sign is displayed if plus pole is
connected to "^"
All display segments available during
operation of the METRAHIT MULTICAL are
activated after the instrument is switched
on.
Measured Quantity
Jump Function
From 0 to 80%
of upper range limit value
From to 50%
of upper range limit value
From 0 to 50%
of upper range limit value
12 mm
30999 steps
GMC-I Messtechnik GmbH
Need help?