Automatic Interval Sequence
Automatic execution of a programmed sequence range is above
all advisable if feed to the signal circuit and scanning of the
peripheral device under test are physically separated.
After entering all parameters for the "automatic interval sequence"
output function (see above) (I
be started and stopped whenever desired, as well as resumed.
GMC-I Messtechnik GmbH
nt
mode
auto
,
=
), the sequence can
Calibration Functions: V – Hz
Example of an Automatic Interval Sequence
U [V]
run up
run dn
hold
LCD:
HOLD CONT
10
7.3
4.6
2
0
5
10
15
20
25
/
➀
➁
Interval parameters: output quantity: U (range of 0 15 V),
start
end
= 2 V,
= 10 V, number of interval steps = 3,
mode
=
auto
Key
1 When 1nt
ready
appears at the display:
Start the sequence by pressing
2 The sequence is stopped by pressing the or the key.
Interval time elapsed thus far is saved as t
3 The sequence is resumed by pressing the key, and
remaining sequence duration t
4 Stop the interval sequence by pressing the
– – Temp – mA
– mA
run up run dn
stop
t [s]
ESC
➂
➃
t1
OUT
ENTER
.
x
= t1 – t
.
y
x
ZERO/SEL
ESC
= 5 s,
key.