7. CALIBRATION
 Turn on the scale and press
 Use the numeric key to enter password
Default password 0000
Press
to confirm
 Display will be shown
Select Local or Remote scale by pressing
Press
to confirm.
 Display will be shown
If necessary, press
Press
to confirm.
 Display will be shown
Ensure the platform is empty and wait for stable indicator.
Press
to confirm.
 Display will be shown
Set weight value will be required
Enter the value by using numeric keys
Press
to confirm.
 Display will be shown
Place the calibration weight on the platform and
wait few seconds for display stable.
Press
to confirm.
 After the calibration scale will start a self test, remove the weight during that time
and display will return to weighing mode.
 Incase display will show any error message or incorrect measurement, repeat the
calibration again.
during the self test.
to select the weighing unit kg or lb.