Owner's Guide
The Coltolux Comfort unit is intended for use in the electromagnetic environment specified below . The customer or the end user of the Coltolux Comfort unit should assure
it is used only in such an environment .
Immunity Test
Electrostatic discharge (ESD)
±8kV Contact
IEC 61000-4-2
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
Electrical fast transient/burst
±2kV for power supply lines
IEC 61000-4-4
100 kHz repetition frequency
Surge
±1kV differential mode (line-line)
IEC 61000-4-5
±2kV common mode
(line-earth)
Voltage dips, short
0 % UT; 0,5 cycle
interruptions and voltage
At 0°, 45°, 90°, 135°,
variations on power supply
180°, 225°, 270° and 315°
input lines
IEC 61000-4-11
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
Power frequency (50/60Hz)
30A/m
magnetic field
50/60 Hz
IEC 61000-4-8
Note UT is the a . c . mains voltage prior to application of the test level .
16
EN/IEC 60601 - 1 -2 2014
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC60601 test level
±8kV Contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
±2kV for power supply lines
100 kHz repetition frequency
±1kV differential mode (line-line)
±2kV common mode (line-earth)
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
30A/m
50/60 Hz
Compliance Level
Intended Electromagnetic Environment
Floors should be wood, concrete or ceramic tile . If
floors are covered with synthetic material, the relative
humidity should be at least 30% .
Mains power quality should be that of a typical
commercial or hospital environment .
Mains power quality should be that of a typical
commercial or hospital environment .
Mains power quality should be that of a typical
commercial or hospital environment . If the user
of the Coltolux Comfort unit requires continued
operation during power mains interruptions, it is
recommended that the Coltolux Comfort unit be
powered from an uninterruptible power supply with
sufficient capacity to run the unit for the maximum
required time of interruption .
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment .