Calibration
Before calibration, the pump must run for 10 minutes MINIMUM.
Previous run must be cleared to calibrate. To clear previous run hold PUMP button until screen
shows four dashes ( - - - - )
1. While the pump is powered on, but not sampling, press both the up and down arrows
simultaneously to enter the Set/Calibrate screen.
NOTE: While in this screen, the calibration will only show on the field calibrator itself.
2. Pressing up or down on the arrows will adjust the flow accordingly on the calibrator.
3. When you have your desired flow, press the "Pump" button to set your flow rate.