All other languages are translations of the original manual. These installation and operating instructions Wear protective gloves. apply to: Tyscor VS 2 Plus REF: 7188200100 Disconnect all power from the unit. Tyscor VS 4 REF: 7188100100 Refer to the accompanying electronic documents.
Important information Safety Fragile, handle with care The unit has been developed and designed in such a way that dangers are effectively ruled out This way up / store and transport in an if used in accordance with the Intended Use. upright position Despite this, the following residual risks can remain:...
Important information General safety information The following groups are not permitted to operate or use a commercially operated unit: – Always comply with the specifications of all – People without the necessary experience and guidelines, laws, and other rules and regula- knowledge tions applicable at the site of operation for the –...
Page 8
Important information If required, the original packaging for the unit can be ordered. The manufacturer and the distributor do not accept liability, even during the war- ranty period, for damage during transpor- tation due to improper packaging. – Only transport the unit in its original packaging. –...
Product description Technical data Electrical data 7188100100 7188200100 Rated voltage 230, 1~ 230, 1~ Mains frequency 50 / 60 50 / 60 Nominal current Rated power Fuses 2x T 10.0 AH / 250 V~ (IEC 60127-2) Type of protection IP 21 Protection class Control connection electrical data Output:...
Page 12
Product description General data Dimensions (H x W x D) 51 x 45 x 69 Weight Noise level * ca. dB(A) 61 / 64 61 / 64 Noise level in accordance with ISO 3746 Network connection LAN technology Ethernet Standard IEEE 802.3u Data rate Mbit/s...
Product description Characteristic curves Fig. 1: Characteristic curves for 7188200100, measured in accordance with ISO 10637 Fig. 2: Characteristic curves for 7188100100, measured in accordance with ISO 10637 7188100021L02 2403V007...
Page 14
Product description Type plate The type plate can be found on the housing lower part. 1 Type plate Evaluation of conformity This device has been subjected to conformity acceptance testing in accordance with the cur- rent relevant European Union guidelines. This equipment conforms to all relevant requirements.
Product description Operation Network connection Mains connection (power plug) Connection to amalgam separator (appliance power inlet) CAN bus Protective strainer Inlet connection Waste water connection Exhaust air connection Orange LED – Ethernet 10 Red LED – fault 11 Green LED – ready for operation 12 Blue LED –...
Product description The VS suction unit is used in "wet" suction systems. The device contains two radial blowers and a sep- aration system. The radial blower and the separation system are each driven by their own motor. Separation system In the separation system the aspirated fluids and the solid particles are separated from the suction air. The separation system has two stages.
Page 17
Product description Eco Stop The Eco Stop function is used to protect the unit if it is operated inadvertently with no flow rate or with a flow rate that is too low. If the unit is operated under these conditions without the manifold signal being actuated in the mean time, the unit will switch off automatically after a pre-defined period of time (it is possible to set this up so that it can be adjusted via the monitoring software).
Page 18
Product description Menu window 1 Title bar / submenus 2 Content area 3 Navigation buttons 4 Scroll button 5 Page indicator 7188100021L02 2403V007...
Page 19
Assembly The following materials must not be used: Assembly – Acrylonitrile-butadiene-styrene (ABS), – Styrene copolymer blends (e.g. SAN + PVC). Requirements Hose materials For waste connections and suction lines only The unit can be installed on the same level as the use the following hose types: surgery room or in a floor below.
Assembly System components Installation type Line layout (minimum requirements) The system components listed below are Flexible – PVC data cable with required or recommended for various procedures shielded cable sheath- or for installation. ing, as used for tele- Rinsing unit communications and IT processing systems It is recommended that the suction system is...
Page 21
Assembly accelerator. The next time suction takes place using the large cannula, the collected fluid is transported in surges and at high speed to the suction system. This ensures automatic cleaning of the suction pipes. 7188100021L02 2403V007...
Assembly Installation The actual connection can vary depending on the chosen installation option. The connection shown is only an example. Installation and routeing of hoses and pipes – Establish connections between the pipe system and the unit using the flexible hoses supplied. This will prevent vibrations from being transmitted to the pipe system.
Page 23
Assembly Installing the group of devi- To install the devices and combine them as a group, use the Installation Kit Tyscor V/VS Tan- dem (see "Optional equipment"). The installation kit contains various parts (pipes, moulded sec- tions, hose clamps etc.) and corresponding installation information required for the installation of two or more devices.
Assembly Network connection CAN bus Purpose of the network connection Purpose of the CAN bus connection The network connection is used to exchange In a grouping of up to three devices, the CAN information or control signals between the unit bus is used to enable mutual controlling and and a software installed on a computer, in order information exchange between the devices.
Page 25
Assembly 1 Mains cable with socket and country-specific mains plug 2 Control line 7188100021L02 2403V007...
Page 26
Assembly PCB (main board) electrical connections X14 X13 X15 X16 Supply voltage 230 V Control voltage output, 24 V DC, control voltage input Voltage supply for the rinsing unit, 24 V DC SD card holder (for Micro SD) CAN bus Service interface Connection for the display Supply voltage for suction motor 1...
Page 27
Assembly Green LED – HV voltage supply +15 V Green LED – HV voltage supply +5V Green LED – HV voltage supply +3.3 V Green LED – temperature indicator for radial blower 1, temperature OK H10 Yellow LED – line to radial blower 1, short circuit or open circuit H11 Red LED –...
Assembly Commissioning Testing suction power after installation In many countries technical medical prod- ucts and electrical devices are subject to The following procedure is recommended in regular checks at set intervals. The owner order to measure the maximum negative pres- must be instructed accordingly.
Page 29
Assembly – Adaptive operation Switches to the start screen – Parallel operation In parallel operation, increased power consumption and increased wear are to Switches to the next higher menu level be expected. Changing to the Administrator user level: Setting the afterrun delay time In the display screen, use the Settings button to switch to the menu screen.
Page 30
Assembly To activate the Eco Stop function, switch from Monitoring the unit via the the display screen to the menu screen by tapping network the Settings button. Then switch to the submenu The following requirements must be met in order via the buttons Parameters and Start/Stop.
Page 31
Assembly Network protocols and ports Port Purpose Service 45123 UDP, Unit recognition and 45124 UDP configuration 1900 UDP Service indicator SSDP / UPnP 502 TCP Device data Event log data Syslog 22 TCP, 23 Diagnosis Telnet, 123 UDP Time The port may vary depending on the configu- ration.
Usage 10.5 Fault Usage Red display flashes – Fault/failure of a radial blower 10 LEDs RED LED illuminates – Failure of both radial blowers – Failure of separation (VS only) 1 Orange LED – network 2 Red LED – fault 3 Green LED –...
Usage 11 Overview on the touch To access the individual menu items, tap the rele- vant button. If there are multiple pages, the arrow screen buttons are used to scroll through the pages. The buttons at the bottom edge of the screen can be used to switch between the different menus.
Usage 12 Disinfection and cleaning Note Important information about the device NOTICE Notice Device malfunctions or damage due Operation of the device restricted to use of incorrect media Guarantee claims may become invalid Fault as a result. Operation of the device interrupted Do not use any foaming agents such ❯...
Usage The following are required for disinfection/clean- ing: ü Non-foaming disinfectant/cleaning agent that is compatible with the materials. ü Unit care system, e.g. OroCup To pre-clean, suck up 2 litres of water with the care system. Aspirate the disinfection/cleaning agent with the care system.
Usage 13 Maintenance All maintenance work must be performed by a qualified expert or by one of our Service Techni- cians. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯ Wear protective equipment when working (e. g. impermeable gloves, protective goggles and ❯...
Troubleshooting Troubleshooting 14 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯...
Page 38
Troubleshooting Error Possible cause Remedy Water leaking from the Membrane valve defective Check the membrane valve at ❯ the waste water connection exhaust air connection and if necessary clean or replace. Foam in turbine due to use of Use non-foaming cleaning ❯...
Page 39
Troubleshooting Error Possible cause Remedy Display does not respond to Display not calibrated Calibrate the display. ❯ touches – Disconnect and reconnect the mains plug. Within 8 seconds of the device being started, start the calibration process with a long press (5 s) on the display.
Troubleshooting 14.2 Error messages Error messages are displayed on the touch screen. If there is a network connection, the mes- sages can be forwarded to the monitoring software. If the device is not connected to the net- work, the messages can be read via a terminal client (e. g. PuTTY). Error Possible cause Remedy...
Page 41
Troubleshooting Error Possible cause Remedy DC bus overvoltage Control error Replace the electronics. * ❯ Mains power supply fault Check the mains connection DC bus undervoltage ❯ and supply voltage. * Machine was disconnected from No action required. ❯ the mains while running Control error Replace the electronics.
Page 42
Troubleshooting Error Possible cause Remedy Eco Stop. Switch start signal Unit has been switched on unin- Check whether all suction ❯ tentionally for too long. hoses are correctly hung up. off and on again to restart Briefly disconnect the unit ❯...
Troubleshooting 15 Transporting the unit WARNING Infection due to contaminated unit Disinfect the unit before transport. ❯ Close all media connections. ❯ Wear protective equipment to avoid any risk of infection (e.g. liquid-tight protective gloves, protective goggles, face mask). Prior to disassembly, clean and disinfect the suction unit and the unit via aspiration of a suitable disinfectant approved by the manu- facturer.
Appendix Appendix 16 Menu structure Setup wizard Language Operating mode Exhaust air filter Finished Start screen Operating mode, vacuum Performance, tem- perature Settings Access level Operator Device information Administrator Service Technician XXXXXX [PIN] Device data Firmware Upgrade Library version PCB_SN File Device usage data Operating hours...
Page 45
Appendix Settings (con- Cluster settings Main unit Parameters tinued) Auxiliary unit 1 Auxiliary unit 2 XX s Lag time Quick Start Start / stop Eco Stop Exhaust air filter Present/not present Control strategy Adaptive operation / parallel operation options List of alarms Alarm information Message history Maintenance...
Appendix 17 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
Appendix 18 Country representatives Country Address Уповноважений представник в Україні: Приватне підприємство “Галіт” вул. 15 квітня, 6Є, с. Байківці, Тернопільський р-н, 47711, Україна тел.: 0800 502 998; UA.TR.099 +38 050 338 10 64 www.galit.te.ua; e-mail: office@galit.te.ua Виробник: Дюрр Дентал ЕсE Хьопфігхаймep Штрассе...
Page 52
Hersteller / Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...