1.5 Notes on EMC and medical 1.7 Incorrect usage products Anyusageaboveandbeyondthatdescribed hereistobeconsideredasincorrectusageof Medicalproductsshouldbetreatedwithrespect theappliance.Themanufacturerisnotliablefor whenitcomestoelectro-magneticcompatibility anydamagesarisingfromincorrectusage.The andspecialsafetymeasuresmustbetaken. operatorbearsallrisksandliabilities. Specialinstructionsconcerningelectro-ma- gneticcompatibilityformedicalproductscan This appliance must not be used in befoundinourspecialleaflet,ordernumber operating theatres or similar rooms, 9000-606-67/30,ortheinformationcanbe or any room where dangers such as foundontheinternet(www.duerr.de)inthe combustion of flammable materials TechnicalDocumentationdownloadpages.
3. Warnings andsymbols 3.2 VistaScan Perio Plus Themodelidentificationplatecanbefoundon 3.1 Installation and therearsideoftheVistaScanPerioPlus. operating Instructions Thefollowingterminologyand/orsymbolsare appliedonthisappliance,inthisappliance,the Intheoperatinginstructionsthefollowingwar- modelidentificationplateandthepowerunit: ningsandsymbolshavebeenused: WarningLaser Information and/or mandatory regu- lations or prohibitions for the preven- tion of personal injury or substantial Warningdangerousvoltage. property damage Specialinstructionsconcerningecono-...
6. Technical Data 6.1 VistaScan Perio Plus Conformity Acceptance Procedure Electrical Details power unit Medicaldevice ....93/42/EWG Voltage(V) ..... .100-240...
VistaScan Perio Plus Always carry the VistaScan Perio Plus by the underside of the housing (Fig. 1). Never carry the VistaScan Perio Plus with the output slot or the cover, since this can damage sensitive components of the VistaScan Perio...
11. Set-up 11.1 Room for Set-up •TheVistaScanPerioPlusshouldbesetupin aclosed,dryandwell-ventilatedroom. •Theroomtemperatureshouldnotfallbelow+ 10°Corexceed+35°C. Danger of short circuit due to build up of condensation. The system can only be put into operation once the appliance has warmed up to room temperature and it is dry. •Installationinapurpose-builtroom,e.g.ina boilerroomorinawetroomisnotpermitted.
Use of the Appliance inactiveside: 14. Image plate Image plates are toxic: for this reason image plates must be placed in a light protection cover before placing into the mouth of a patient. In the case of a patient swallowing an image plate or a part of one then a doctor must be consulted immedi- ately and the image plate removed.
18. Hygienic covers •provideaddedprotectiontothepatientsfrom infectionbycontaminatedlightprotection coversandimageplates. •alsoprotectstheimageplatefromcontamina- tionanddirt. •protectthedevicefromcontaminationfrom contaminatedlightprotectioncovers. •isavailableindifferentsizesjustlikeimage plates.Choosethemostappropriatesize. Loss of image data Image data on the image plate will be lost when exposed to light. Never handle exposed image plates wi- thout the light protection cover. The hygienic cover does not protect the plates from light exposure.
20.2 X-ray exposures Dangerofinfection! Whenplacingtheimageplateintothe patient'smouthalwayswearprotective gloves. Theimageplatesaremarkedwitha symbol .Thesymbolisvisibleinthe X-rayexposure,andmakesorientation duringdiagnosiseasier. Theimageplateshouldbeplacedinthe patent'smouthsothatthesymbolwill notaffectthediagnosticusefulnessof theX-rayexposure. •Placetheimageplateinsidethelightprotec- tioncoverasnormalintothepatient'smouth. Thesilverpointontheinactivesidemustface theocclusalarea.Theactivesideoftheimage platemustpointinthedirectionoftheX-ray tube. Damage to the image plate Only use handling systems which will not cause any damage to either the light protection cover or the image plates.
•Laytheimageplatedowninthelightprotec- tioncoverinthedisinfectantwipe. •Takeofftheprotectivegloves,disinfectthe hands(e.g.withDürrDentalHD410)and clean(e.g.withDürrDentalHD430,HD435). Damage to the VistaScan Perio Plus Before handling the image plates, hands should be completely clean and free of any traces of powder from protective gloves. This powder can damage the optical scanning unit of VistaScan Perio Plus during the scanning procedure of the image plate.
Page 28
•Pushthecassette(9b)intothecassettecover (9c). •Flipthecover(9a)ofthefoilcassetteopen untilitlatchesonce. •Placetheopenedlightprotectioncovertight againstgap1orgap2. Theblack(inactive)sideoftheimageplate mustbevisible. •Slidetheimageplateoutofthelightpro- tectioncoverandimmediatelyintothefoil cassetteuntilitisfirmlyinposition. •Closethecover(9a)onthefoilcassette. Damage to the image plate: The surface of the image plate can be sealed off if the foil cassette is pushed in improperly. The light protection cover must therefore lay firmly against the cover of the foil cassette.
In order to avoid confusing these X-rays from those of other patients only X-ray exposures for this (one) patient should be scanned at one Transport slot status LEDs time. VistaScan Perio Plus under DBSWIN provides additional data, Problem e.g. patient number, room, etc. on Imageplatecurrentlybeing the display.
Page 30
The following actions are possible on both transport slots at the same time: StatusLEDisgreen: •Insertthefoilcassettewiththeimageplateon theinputunit(Fig.27). •Useryourfingerstoslidethecassetteoutof thecassettecoverintothetransportslot(Fig. 28). StatusLEDilluminatesinyellowandgreen: •Removethecassettecover. •Insertthesecondfoilcassettewiththeimage plateontheinputunit. •Useryourfingerstoslidethecassetteoutof thecassettecoverintothetransportslot. StatusLEDisyellow: •Donotinsertanymorecassettes. Theprogressofthereadprocedureisshownin thepreviewwindowonthemonitorandonthe display. •TheX-rayexposurewillbesavedinDBSWIN.
21.3. Foil cassettes Damage to foil cassettes Do not put foil cassettes in the autoclave! Visible dirt must be removed before the foil cassette is used again! •Alternatively,useadisinfectantspray (e.g.DürrFD322orFD333)onasoft, fluff-freeclothinsteadofadisinfectantwipe fordisinfecting. Alternativelythefoilcassettescanbeim- mersedinabathcontaininganon-aggressive disinfectant(e.g.DürrID212forte). •Beforereusage,thecassettemustbe completelydry. 21.4. Storage box with image plate storage tray Storage box Cleananddisinfectsurfaceswithcontamination...
21.5 VistaScan Perio Plus Damage to the VistaScan Perio Plus Fluids must not be allowed to enter the inner parts of the appliance, the- refore do not use spray disinfection or cleaning solutions on the Vista- Scan Perio Plus! Appliance surfaces...
Trouble-shooting 23. Tips for Operators and Technicians Repairs may only be carried out by suitably qualified personnel or our own service technicians. 23.1 VistaScan Perio Plus Problem Probable cause Solution 1. The VistaScan Perio •Nosupplyvoltage •Checkthemainscableandall Plus does not switch connections •Checkthemainfuseboxto...
Page 37
Problem Probable cause Solution 7. X-ray exposure •Imageplatehasbeenputin •Rescantheimageplateas does not appear on cassetteupsidedownandthe quicklyaspossible the monitor after inactivesidescanned scanning •Imagedataontheimageplate •Imagedataontheplatemustbe hasbeenerased,e.g.byambient scannedasquicklyaspossible light afterextraction •Problemwiththehardware •InformyourServiceTechnician •X-raydeviceisfaulty •InformyourServiceTechnician •Noimageinformationonthe •Exposetheimageplate imageplate 8. X-ray too dark / •Incorrectbrightness/contrast •SetthebrightnessoftheX-ray light settinginthesoftware imageusingthesoftware...
Page 38
Problem Probable cause Solution 13. VistaScan Perio •ThesubnetmaskofthePCand • C heckthesubnetmasksand Plus does not ap- VistaScanPerioPlusdonot adaptasrequired pear in the options match list in VistaScan- Config. Connection is not possible however! 14. Error message •VistaScanPerioPlushasacon- •SetVistaScan/VistaScanPerio appears on starting flictwithanotherappliance Plusto"DemowithoutScanner" X-ray module in in"VistaScanConfig".
Page 39
Problem Probable cause Solution 19. Error message •Aconnectionwasbrokenoffwhile "E2490" thesoftwarewasaddressingthe VistaScanPerioPlus. 20. Cassette is not •filmcassettenotcorrectlyinserted •Placethefilmcassettecorrectly drawn into the •Othercassettesareintheinput •Waituntilinputbufferisempty. transport arc buffer,statusLEDilluminatedin StatusLEDisilluminatedin orange green: •Whenthetransportmechanism •Seesection20.4Readingimage isnotactivated,VistaScanPerio data Plusmaynotbereadyforoperati- •Cassettecoverisnotonthe •Putthecassettecoveronand cassette positionagain 21. Cassette is not •Cassetteremainsstuckinthe •InformyourServiceTechnician transported into...