4.4 Cleaning

Precleaning
All individual parts must be precleaned immediately after use.
EQUIPMENT:
– Drinking water temperature of about 15 °C
PROCEDURE:
• Rinse all parts used for 2 minutes in running drinking water.
Manual cleaning
EQUIPMENT:
– Drinking water temperature of about 40 °C
– Standard commercial washing-up liquid
– Receptacle with at least 3 l capacity
Procedure
• Add about 1 teaspoonful washing-up liquid to 3 l warm
drinking water.
• Place all the parts in the washing-up water.
Application time: 5 minutes
• Occasionally move the parts back and forth.
• In case of visible soiling, use a medium-soft brush
(e.g. a toothbrush) which is used exclusively for this purpose.
RINSING:
• Rinse off all parts thoroughly in running drinking water at about 15 °C for 3 minutes.
DRYING:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean and absorbent surface and allow them to dry completely.
If the pressure gauge connector line is visibly soiled, replace it.