REPROCESSING

The product components must be cleaned thoroughly immedi-
ately after each use and disinfected at least once a day.
If the nebuliser is used in professional environments, follow
the information on reprocessing included in the appendix at the
end of these instructions for use.

4.1 Preparation

• Detach the tube from the nebuliser.
• Detach the mouthpiece from the nebuliser.
• Make sure that all residual volume is removed from the
nebuliser.
• Dismantle the nebuliser into its individual parts.
• Carefully pull the blue exhala-
tion valve out of the slot in the
mouthpiece. The exhalation
valve must still be attached to
the mouthpiece.

4.2 Cleaning

The connection tube cannot be cleaned or disinfected. Re-
garding care of the connection tube, follow the instructions
in the corresponding section [see: Care of the connection
tube, page 22].
• Briefly rinse all parts used in running drinking water
beforehand.
• Place all disassembled com-
ponents in warm drinking water
with a little dishwashing liquid
for about 5 min.
• Rinse off all parts thoroughly in drinking water.
• Shake the water out of all of the parts.