Advertisement
The PARI inhalation system consists of a PARI compressor, a PARI nebuliser and PARI accessories. The system is used to treat the lower airways.
This product can be used in a home environment, as well as in professional health institutions. Aside from the compressor, if used in a home environment, the product components may only be used by a single patient (no patient change). In a professional environment, the device can be used with different patients as long as the corresponding hygiene reprocessing measures are complied with.
This product must be used only by individuals who understand the contents of the instructions for use and are able to use the product safely.
Individuals in the following groups must be supervised by a person who is responsible for their safety:
If the patient is not able to use this product safely on their own, then the treatment must be carried out by the responsible person.
This PARI product is suitable only for patients who are able to breathe by themselves and are conscious.
The frequency and duration of use is determined by professional medical staff [1]according to the individual needs of the patient.
Compressor
The purpose of the PARI compressor is to generate compressed air for operating a PARI nebuliser.
The PARI compressor must be used only for PARI nebulisers. It can be operated by the patient themselves and must only be used indoors.
Nebuliser
The PARI LC SPRINT nebuliser generates an aerosol[2] for inhalation for the therapy of the lower airways.
Together with a PARI compressor or the PARI CENTRAL and with PARI accessories, the nebuliser forms an inhalation system.
The nebuliser is suitable for use in treating patients in all age groups.
Only solutions and suspensions that are approved for use in nebuliser treatment may be used.
The nebuliser must only be connected with a PARI compressor or with a central gas supply system. The PARI CENTRAL is intended for the connection with the central gas supply system.
An application takes approximately 5 to 10 minutes (depending on the quantity of fluid), but in any case no more than 20 minutes.
[1] ) Professional medical staff: Doctors, pharmacists, and physiotherapists.
[2] ) Aerosol: Small particles of solid, liquid or mixed composition (fine "mist") suspended in gases or air.
Mask
The PARI mask soft is an accessory for inhalation treatment. That enables inhalation of aerosol2 through the mouth and nose.
The different mask sizes are suitable for treating patients in the following age groups:
The specified ages are approximate. The actual size of the mask depends on the size and shape of the person's face.
The mask must be used only with PARI nebulisers.
Indication
For treatment of diseases of the lower airways.
Masks
For patients who cannot inhale using a mouthpiece, or if inhalation via mask is preferred. The mask forms a system with a nebuliser. The indication for this system is the same as the indication for the nebuliser used.
Contraindication
Compressor, nebuliser and mask
There are no contraindications known to PARI GmbH.
The following symbols can be found on the product and/or the packaging:
![]() | Medical device |
![]() | Unique Device Identifier (UDI) |
![]() | Legal manufacturer |
![]() | Date of manufacture |
![]() | Serial number |
![]() | Item no. |
![]() | Production batch number, lot number |
![]() | This product conforms to the EU Medical Device Regulation 2017/745. |
![]() | Consult instructions for use |
![]() | The device is drip-proof (degree of protection as per IEC 60529/EN 60529). |
![]() | Humidity limit |
![]() | Atmospheric pressure limit |
![]() | Degree of protection of the application component: Type BF |
![]() | Protection class II appliance |
![]() | Temperature limit |
![]() | Alternating current |
![]() | The medical device was distributed commercially after 13 August 2005. The product must not be disposed of with normal domestic waste. The symbol of the refuse bin with a cross through it indicates that it must be disposed of separately. |
![]() | On/Off |
![]() | PARI BOY compressor |
![]() | Air filter for compressor type 130 |
![]() | Tubing adapter |
![]() | PARI adult mask soft |
![]() | Elastic band |
![]() | Mask stabiliser |
![]() | PARI LC SPRINT nebuliser with nozzle attachment |
![]() | Connection tubing |
![]() | Mouthpiece with exhalation valve |
The present instructions for use contain important information, safety instructions and precautionary measures. The user must follow these in order to guarantee safe operation of this PARI product.
This PARI product must be used only as described in these instructions for use. The instructions for use of the inhalation solution used must also be followed.
Labelling and classification of warning instructions
In these instructions for use, safety-critical warnings are categorised according to the following hazard levels:
DANGER indicates a hazardous situation which will lead to very severe injuries or death if it is not avoided.
WARNING indicates a hazardous situation which can lead to very severe injuries or death if it is not avoided.
CAUTION indicates a hazardous situation which can lead to mild or moderate injuries if it is not avoided.
NOTE
NOTE indicates a hazardous situation which can lead to material damage if it is not avoided.
General
If non-approved solutions or suspensions are used for nebulisation, then nebuliser aerosol characteristics may differ from the information provided by the manufacturer.
This product is not suitable for use in an anaesthetic breathing system or a ventilator breathing system.
Tracheotomised patients cannot inhale using a mouthpiece. To perform inhalation therapy, they require specific equipment. In this case, please contact your doctor for further information.
Skin care products containing oils or fats can damage the soft plastic components. The patient should refrain from using skin care products of this kind while using the device.
If your health condition does not improve or it even worsens as a result of the treatment, seek professional medical advice.[3]
[3] ) Professional medical staff: Doctors, pharmacists, and physiotherapists.
Life-threatening situation from electrocution
Life-threatening situation from electrocution The compressor is an electrical device powered by mains voltage. It has been designed so that no live parts are accessible. However, in unfavourable ambient conditions, or if the compressor or power cord is damaged, this protection may no longer be provided. There may then be a risk of contact with live parts. This in turn may lead to an electric shock.
|
Danger of a device defect
To avoid a device defect, comply with the following instructions:
Treatment of babies, infants, and anyone who requires assistance
|
Life-threatening situation from strangulation For individuals who are not able to perform the therapy session without assistance or cannot appreciate the hazards, the risk of injury is greater e.g., strangulation with the power cord or the connection tubing. Such individuals include, for example, babies, children, and people with limited capabilities.
|
Hazard due to small parts which can be swallowed
The product contains small parts. Small parts can block the airways and lead to a choking hazard. Keep all components of the product out of the reach of babies and infants at all times.
Impairment of treatment due to electromagnetic interference
Use only original spare parts and original accessories from PARI. The use of third-party products can result in increased emissions of electromagnetic interference or reduced interference resistance of the PARI compressor.
Impairment of the therapy by disregarding the reprocessing instructions
Impairment of the therapy
Exceeding the reprocessing instructions can result in damage to the product. This can impair the therapy.
Falling short of the instructions can result in insufficient reprocessing. This can increase the risk of infection.
Reporting serious incidents
Report serious adverse incidents to the manufacturer and to the competent authority.
Components
Please refer to the package for information on the supplied components.
1 | Compressor | |
2 | Nebuliser | |
2a | Nebuliser upper part | |
2b | Nozzle attachment | |
2c | Nebuliser lower part | |
2d | Mouthpiece (with exhalation valve) | |
2e | Tubing adapter | |
2f | Connection tubing | |
3 | Mask | |
3a | Elastic band |
The compressor includes the following working parts:
1 | Carrying handle | |
2 | Power cord[4] (connected inseparably to the compressor) | |
2a | Cable holder | |
3 | Ventilation slits | |
4 | Holder for nebuliser | |
5 | Identification label (bottom of device) | |
6 | Compressed air connection | |
7 | Air filter | |
8 | On/off switch |
[4] ) The power plug type is country-specific. The figure shows the europlug (type "C").
Inhalation therapy (compressor with nebuliser and mouthpiece or mask)
The compressor supplies the nebuliser with compressed air. When compressed air is supplied, the nebuliser generates an aerosol from the liquid, for example the medication with which it is filled. This aerosol is breathed into the lungs through the mouthpiece or optionally a mask.
In combination with a mouthpiece or a suitable mask, the PARI LC SPRINT nebuliser (Type 023) is a suitable therapy for patients in all age groups. The nozzle attachments described below are particularly suitable for the specified age groups.
The size of the aerosol droplets is determined by the nozzle inserts. The smaller the droplets are, the farther they can penetrate into the deeper and smaller regions of the lung:
Mask
The mask is an accessory for PARI inhalation systems.
The mask makes it possible to inhale aerosol through the mouth and the nose.
The patient can breathe out through the aperture or the exhalation valve at the bottom end of the mask without having to take the mask off.
The PARI child and adult mask soft can be fixed to the face using the elastic band. The elastic band is attached to the loops on the side of the mask.
Only in professional healthcare institutions is the mask stabiliser inserted into the mask during the sterilisation process, to maintain the shape of the mask.
The individual product components are made from the following materials:
Nebuliser
Product component | Material |
Nebuliser upper part | Polypropylene, thermoplastic elastomer |
Nozzle attachment | Polypropylene |
Nebuliser lower part | Polypropylene, thermoplastic elastomer |
Mouthpiece (with exhalation valve) | Polypropylene, thermoplastic elastomer |
Polyvinyl chloride (PVC) connection tubing | Polyvinyl chloride |
Tubing adapter | Polyamide |
Tubing connector | Thermoplastic elastomer |
Mask
Product component | Material |
PARI mask soft | Polypropylene, thermoplastic elastomer |
Elastic band | Synthetic rubber |
The compressor is maintenance free.
The individual product components have the following expected lifetimes:
Product component | Service life |
Compressor | Approximately 1,000 operating hours (this corresponds to max. 5 years) If the compressor is still in use after this time, have it tested. To do this, contact the manufacturer or distributor. |
Nebuliser, PARI mask soft, tubing adapter, connection tubing und accessory | Home environment [see: Limits of reprocessing in a home environment,] |
Nebuliser, PARI mask soft, tubing adapter, connection tubing und accessory | Professional environment [see: Limits of reprocessing in a professional healthcare institution] |
When the expected operating life has been reached, replace the affected component. Replacement sets or PARI Year Packs (a nebuliser with connection tubing and an air filter for a compressor) are available.
All the steps described below must be carried out properly.
If the nebuliser is to be connected to a central medical gas supply system via a PARI CENTRAL, the instructions for use of the PARI CENTRAL must be followed.
Do not use product components unless they have been thoroughly cleaned and dried. Wash your hands thoroughly before every use. You must perform cleaning and disinfection before using the device for the first time
Danger of acute respiratory distress when the nebuliser is operated using oxygen When the nebuliser is used with the PARI CENTRAL O2, acute respiratory distress can arise through an increased level of carbon dioxide in the blood.
COPD patients with exacerbation may develop acute respiratory distress as a result of treatment which uses oxygen.
The space in which the compressor is operated must satisfy certain conditions [see: During operation]. Also note the following warnings before setting the compressor up:
Risk of fire due to a short circuit
A short circuit in the compressor can cause a fire. In order to reduce the risk of fire in such an event, follow the instructions below:
Impairment to quality of treatment caused by electromagnetic interference Electrical devices can cause electromagnetic interference. Interference can impair the function of the devices and thus also the effectiveness of the treatment.
Danger of injury from falling compressor
A compressor placed in a poor position poses a risk of injury.
NOTE
Danger of device fault caused by dust
If the compressor is operated in a very dusty atmosphere, dust may collect inside the housing. This may cause a fault in the device.
Set the compressor up as follows:
Assembling the nebuliser
Risk of impaired treatment
Damaged components and/or an incorrectly assembled nebuliser may impair functioning of the nebuliser and thus treatment as well.
Preparing the inhalation therapy
USING THE MOUTHPIECE
USING THE MASK
Filling the nebuliser
NOTE
Nebuliser lid might break off
If the cap is twisted in the wrong direction, it may break off. The nebuliser will then be unusable and irreparable.
If several inhalation solutions are to be used one after the other:
All the safety instructions and warnings in these instructions for use must have been read and understood before any treatment is carried out.
Always hold the nebuliser upright during treatment.
Information about the PIF-Control System:
The PARI PIF-Control System in the nebuliser upper section is designed to help the patient learn a slow, controlled inhalation technique. This improves uptake of the active agent in the lower airways.
If the patient breathes in too quickly, the inflow of air is reduced, and this increases the resistance when inhaling.
If you feel increased resistance when inhaling during the treatment, proceed as follows:
Inhalation therapy
INHALING WITH THE MOUTHPIECE
Some residual fluid will remain in the nebuliser after the end of the treatment.
INHALING WITH A MASK
Impaired treatment due to escaping aerosol
If the mask does not form a seal on the face, aerosol may escape. This may result in medication underdosage.
With PARI child or adult mask soft
Some residual fluid will remain in the nebuliser after the end of the treatment.
Complete disconnection from the mains is only certain when the power plug has been unplugged from the socket.
After ending the treatment, store the compressor as described below:
When putting the compressor away, unplug the power plug from the socket. Electrical devices that remain plugged into the power supply present a potential hazard source.
Risk of infection due to cross-contamination in the case of a change in patients
If a product is used for more than one patient, there is a risk that germs may be transmitted from one patient to the next.
Risk of infection due to growth of bacteria
Inadequate disinfection encourages the growth of bacteria and thus increases the risk of infection.
Risk of infection by residual germs
If there is dirt on the parts, germs capable of reproduction may remain despite the sterilisation process. As a result, there is a danger of infection.
Risk of infection due to moisture
Moisture encourages the growth of bacteria.
NOTE
Risk of damage to plastic parts!
Plastic melts if it comes into contact with hot surfaces.
Always use drinking water for reprocessing in a home environment.
In professional health care institutions, drinking water is sufficient for precleaning. For all other reprocessing steps and processes, use deionised water with a low microbiological load (at least drinking water quality).
Please observe the instructions for use for the chemicals used. Wash your hands thoroughly before every reprocessing.
Perform the reprocessing steps as per the specified reprocessing cycles.
Reprocessing cycles in a home environment
Nebuliser components |
|
Mask components |
|
Elastic band | Cleaning visible dirt |
Compressor housing |
|
Air filter | Replace after 200 operating hours (approx. 1 year)[6] |
Reprocessing cycles in professional healthcare institution
SINGLE PATIENT USE
Nebuliser components |
|
Mask components |
|
Elastic band | Cleaning visible dirt |
Compressor housing | Cleaning visible dirt |
Air filter | Replace after 200 operating hours (approx. 1 year)6 |
BEFORE A CHANGE OF PATIENTS
Nebuliser components |
|
Mask components |
|
Connection tubing | Mechanical cleaning with disinfection |
Elastic band | Replacing the elastic band |
Compressor housing |
|
Air filter | Replace after 200 operating hours (approx. 1 year)6 |
[5] ) For the therapy of patients at risk, disinfect the individual parts once per day. Take note of the additional information for patients at risk [see: Reprocessing for patients at risk].
[6] ) Air filters are included in every PARI Year Pack.
Reprocessing for patients at risk
Danger of infection for patients at risk
For patients at risk, airway infections represent an increased risk of deterioration of general health, because these patients are particularly vulnerable to residual germs. Patients at risk include cystic fibrosis patients, patients with immunosuppression or immunodeficiency, and vulnerable patient groups.
Limits of reprocessing in a home environment
Nebuliser and accessories, disinfection | 300 processing cycles, max. 1 year |
Mask, disinfection | 300 processing cycles, max. 1 year |
Limits of reprocessing in a professional healthcare institution
Nebuliser and accessories, disinfection | 300 reprocessing cycles, max. 1 year |
Nebuliser and accessories, sterilisation | 100 processing cycles, max. 1 year |
Connection tubing | 50 reprocessing cycles, max. 1 year |
Mask, disinfection | 300 processing cycles, max. 1 year |
Mask, sterilisation | 100 processing cycles, max. 1 year |
Procedure | ||
Compressor |
| |
Nebuliser |
| |
• Carefully pull the blue exhalation valve out of the slot in the mouthpiece. The exhalation valve must still be attached to the mouthpiece. | ![]() | |
Mask |
|
The following products can be cleaned, disinfected and sterilised according to the procedure described below:
The connection tubing and elastic band of the mask must be treated separately.
Procedure | |||
Step 1: Precleaning | All individual parts must be precleaned immediately after use. EQUIPMENT:
PROCEDURE:
| ||
Home environment | Step 2: Cleaning | EQUIPMENT:
| |
PROCEDURE:
| ![]() | ||
RINSING:
Cleaning the elastic band
|
Procedure | ||
Home environment | Step 3: Desinfection |
EQUIPMENT:
PROCEDURE:
|
EQUIPMENT:
PROCEDURE: | ||
In professional health care institutions | Step 2: Cleaning and disinfection | EQUIPMENT:
PROCEDURE:
|
Procedure | |||
In professional health care institutions | Step 3:Sterilisation | For sterilisation, use a process compliant with the ISO 17665 series of standards. PROCEDURE:
| |
Information on sterilising a PARI mask soft: Always use the corresponding mask stabiliser when sterilising this mask type, because otherwise the mask may lose its shape under the effects of high temperatures. | ![]() | ||
| |||
Drying | Ensure that there is no residual moisture remaining in the components. If necessary:
| ||
Visual Inspection | Inspect all product components after each cleaning, disinfection or, where applicable, sterilisation. Replace defective, deformed or seriously discoloured parts. | ||
Storage | Store this product as described below:
|
|
Life-threatening situation from electrocution Liquids can conduct electricity, thereby posing a risk of electric shock.
|
NOTE
Danger of device fault due to liquid penetration
If liquids get into the interior of the compressor, this may cause a fault in the device.
Procedure | ||
Step 1: Cleaning | Wipe the outer surface of the housing with a clean, damp cloth. | ![]() |
Step 2: Desinfection | For disinfection, use a standard, alcohol-based disinfectant (e.g. isopropanol). When applying and dosing the disinfectant, it is essential to follow the instructions for use for the product.
The procedure was validated in Europe using a disinfectant suitable for use on plastics as per the DGHM and/or the VAH list: Propanol/Isopropanol | |
Visual Inspection |
The air filter must be checked at regular intervals (after every 10th use). If it is discoloured brown or grey, or if it is damp or clogged, it must be replaced. The air filter cannot be cleaned and then reused! | |
Removing the air filter
| ![]() | |
Replacing the air filter
| ||
Storage |
| ![]() |
|
Dry the connection tubing after each use.
The maximum operating life of the connection tubing is 1 year.
Procedure | ||
In professional health care institutions (when changing between patioents) | Mechanical cleaning and desinfection | EQUIPMENT:
PROCEDURE:
|
Drying |
| |
Visual Inspection |
| |
Storage |
|
Further validated processes for reprocessing
The instructions provided were validated by PARI and were found to be suitable for preparing your medical device for its reuse.
Further validated processes for reprocessing:
https://www.pari.com/fileadmin/041D0624_Professional_healthcare_institution_Validated_Reprocessing_Methods.pdf
Ensure that the reprocessing actually performed by your personnel on your equipment with the chemicals used achieves the desired results. For this, validation and routine monitoring of the process are usually required. In particular, if you have to deviate from our validated process, ensure that the reprocessing method selected by you is appropriately effective, and that potential adverse effects are assessed.
The compressor must be repaired only by PARI GmbH Technical Service or a service location expressly authorised to do so by PARI GmbH. If the compressor is opened or manipulated by anyone else, all claims under the warranty shall be void. In these cases, PARI GmbH will accept no liability.
Contact the manufacturer or distributor:
Fault | Possible cause | Remedy |
The compressor does not start up | The power plug is not plugged into the mains socket correctly. | Check whether the power plug is plugged into the socket correctly. |
The supply voltage is not suitable for running the compressor. | Check whether the local supply voltage matches the voltage data marked on the compressor identification label. | |
No aerosol is coming out of the nebuliser | The nebuliser nozzle is blocked. | Clean the nebuliser. |
The connection tubing is not connected properly. | Check that the tubing connectors are connected firmly to the compressor and the nebuliser. | |
The connection tubing is leaking. | Replace the connection tubing. |
General compressor data
Supply voltage | 220 – 240 V |
Mains frequency | 50 Hz |
Power consumption | 0.95 A |
Housing dimensions (W × H × D) | 18.5 cm × 13.0 cm × 15.0 cm |
Weight | 1.7 kg |
Pressure[7] | 1.6 bar |
Compressor flow7 | 5.0 l/min. |
Sound pressure level | 54 dB(A) |
Classification as per IEC 60601-1/EN 60601-1
Type of electric shock protection | Protection class II |
Degree of protection from electric shock from the application component (nebuliser) | Type BF |
Degree of protection against penetration by water and solid materials, as per IEC 60529/N 60529 | IP 21 |
Degree of protection when used in the presence of flammable mixtures of anaesthetics with air, with oxygen, or with nitrous oxide | No protection |
Operating mode | Continuous operation |
Electromagnetic compatibility
Electrical medical equipment is subject to special precautionary measures with regard to electromagnetic compatibility (EMC). Such equipment must be installed and operated only in accordance with the electromagnetic compatibility instructions.
Portable and mobile high-frequency communication devices can disrupt electrical medical equipment. Using accessories, converters and power cords other than those specified (with the exception of converters and power cords that the manufacturer of the medical electrical device sells as spare parts for internal components) can result in higher emission levels or reduce the device's resistance to interference.
The device must not be operated directly beside or on top of other devices. If the medical electrical device must be operated beside or on top of other devices, then it must be monitored to ensure that it is operating properly in the arrangement used.
On request, technical data on electromagnetic compatibility (EMC information) is available in table format from the manufacturer or distributor, or on the website [see: Links]
Ambient conditions
DURING OPERATION
Ambient temperature | +10°C to +40°C |
Relative humidity | 30% to 75% (non-condensing) |
Atmospheric pressure | 700 hPa to 1,060 hPa |
The compressor is intended for operation in all healthcare situations. It is not permitted to operate it in trains, motor vehicles or aeroplanes.
Use of the compressor in professional healthcare facilities is limited to the inpatient wards and the intensive care unit. Use of the compressor in areas with elevated magnetic or electrical radiation (e.g. close to an MRI scanner) is not permitted.
TRANSPORT AND STORAGE BETWEEN USES
Minimum ambient temperature (without monitoring of relative humidity) | -25°C |
Maximum ambient temperature (with relative humidity of up to 93%, non-condensing) | +70°C |
Humidity | max. 93% |
Atmospheric pressure | 500 hPA to 1,060 hPA |
General nebuliser data
Size[8] | 10 cm × 10 cm × 4 cm |
Weight8 | 31 g to 33 g |
Operating gases | Air, oxygen |
Minimum compressor flow | 3.0 l/min. |
Minimum operating pressure | 0.5 bar / 50 kPa |
Maximum compressor flow | 6.0 l/min. |
Maximum operating pressure | 2.0 bar / 200 kPa |
Minimum fill volume | 2 ml |
Maximum fill volume | 8 ml |
Aerosol data according to ISO 27427
The aerosol characteristics presented in these instructions for use were determined in accordance with ISO 27427 using 2 ml Salbutamol fill volume. If other solutions or suspensions are used for nebulisation, the aerosol characteristics may differ from the values shown (particularly if they have greater viscosity).
The following data is based on tests according to a standard which takes adult breathing patterns as a basis. Therefore, these figures will probably differ from corresponding figures that were calculated for populations of children and infants.
Nozzle insert (blue) | Minimum compressor flow (3 l/min – 0.6 bar) | Nominal compressor flow (5 l/min – 1.6 bar)[9] | Maximum compressor flow (6 l/min – 1.9 bar) |
MMAD [µm][10] | 4.7 | 3.8 | 3.3 |
GSD[11] | 2.19 | 2.24 | 2.70 |
Respirable fraction [% < 5 µm] | 52.3 | 61.9 | 60.5 |
Aerosol fraction [% < 2 µm] | 13.3 | 22.1 | 29.4 |
Aerosol fraction [% > 2 µm < 5 µm] | 39.0 | 39.8 | 31.2 |
Aerosol fraction [% > 5 µm] | 47.7 | 38.1 | 39.5 |
Aerosol output [ml] | 0.35 | 0.41 | 0.38 |
Aerosol output rate [ml/ min] | 0.07 | 0.16 | 0.18 |
Residual volume [ml] (gravimetric) | 1.16 | 1.16 | 1.10 |
Percentage of fill volume emitted per minute [%/min] | 3.3 | 8.0 | 9.2 |
[7] ) Towards nebuliser nozzle (Ø 0.48 mm).
[8] ) Without mouthpiece and mask; unfilled.
[9] ) Operation with PARI BOY compressor (Type 130).
[10] ) MMAD = Mass Median Aerodynamic Diameter
[11] ) GSD = Geometric Standard Deviation
Compressor
This product falls within the scope of the European Council Directive on Waste Electrical and Electronic Equipment (WEEE)[12]. Accordingly, this product must not be disposed of with domestic waste. The disposal regulations applying in the respective countries must be complied with (e.g. disposal by local authorities or distributors). Materials recycling helps to reduce the consumption of raw materials and to protect the environment.
Nebuliser and mask
All product components may be disposed of with normal domestic waste. The country-specific disposal regulations must be observed.
[12] ) Directive 2012/19/EU of the EUROPEAN PARLIAMENT AND THE EUROPEAN COUNCIL of July 4, 2012 on waste electrical and electronic equipment.
Terms and conditions of warranty: https://www.pari.com/int/warranty-conditions
Technical data regarding electromagnetic compatibility:
https://www.pari.com/fileadmin/041D0623-Electromagnetic-compatibility-EMV.pdf
PARI inhalation systems in aircraft:
https://www.pari.com/fileadmin/041D0625_Airplane_Certificate_Jet_nebuliser.pdf
Contact
Email: info@pari.de
Tel.: +49 (0)8151-279 220 (international)
+49 (0)8151-279 279 (German)
Disclaimer
These instructions for use describe the components of PARI products and optional accessories. For this reason, these instructions for use also describe and illustrate features not present in your PARI product because they are, for instance, country-specific and/or optional. When using the systems, products and functions, the applicable country-specific regulations must be observed.
Warranty
The PARI compressor comes with a 4-year warranty. The warranty period commences on the date of purchase.
PARI GmbH
Spezialisten für effektive Inhalation
Moosstraße 3
82319 Starnberg • GERMANY
info@pari.de
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Need help?
Do you have a question about the BOY Classic and is the answer not in the manual?
Questions and answers