Tektronix 2247A Operator's Manual page 91

Portable oscilloscope
Hide thumbs Also See for 2247A:
Table of Contents

Advertisement

*
Operators
Familiarization
Self Cal Measurements
This selection does a self-characterization of the vertical channels 1 and 2
and the horizontal timing. The SELF
CAL MEASUREMENTS
routine stores
calibration constants
that set the accuracy
of the Internal measurement
system.
Voltage
or tlme measurements
can be calibrated separately
by
selecting the SELF CAL Item from the VOLTMETER
or TIME Measurement
Menus.
NOTE
CALIBRATE
MEASUREMENTS
or
SELF
CAL
can
be
performed anytime after a 20-minute
warmup
to ensure the
accuracy stated in Section
6.
Internal Settings
Menu
The INTERNAL SETTINGS MENU
consists of MAKE FACTORY SETTINGS,
LOAD
STORE/RECALL
STORED
SETUPS,
and
ADJUST
VERTICAL
OUTPUT.
All
INTERNAL
SETTINGS
menu
items,
except
ADJUST
VERTICAL
OUTPUT,
are for use by the operator.
To return to the main
SERVICE
MENU,
press
the
up-arrow
key
to
underline
INTERNAL
SETTINGS
MENU
and
press
the
button
again
to
display
the
SERVICE
MENU.
To
return
to
the
oscilloscope
mode,
select
QUIT
from
the
INTERNAL
SETTINGS
MENU
or
SERVICE
MENU,
or
press
the
CLEAR
DISPLAY button.
MAKE
FACTORY
SETTINGS—Sets
the
front-panel
controls
and
menu
configurations
as described
In Appendix
B.
INTERNAL SETTINGS MENU/
MAKE FACTORY SETTINGS:
T MENU
LOAD STORE/RECALL SETUPS
t
ADJUST VERTICAL OUTPUT
quit
Internal
Settings
Menu
2247A Operators
3-39