Controls, Connectors,
and Indicators
In addition,
applied
signals
must
be
large
enough
to insure
a stable
trigger In NOISE REJECT at 2 mV/dlv and be within the dynamic range
of the vertical and trigger system
at 5 V/div.
The
Input signal must be
DC
or perlodic within the frequency
range
of 50 Hz to 100 MHz.
Auto-setup
actlon
for
each
front-panel
control
is
shown
in
the
Appendix of this manual In Table A-1.
In addition to the actlon shown
In Table
A-1,
any
selected
measurement
is re-initialized
as if it were
manually
selected.
NOTE
MIN SETUP
as described below is only available on instru-
ments Serial Numbered
B020100 or above.
MIN
SETUP—Pushing
and holding the Auto-Setup
button until Instru-
ment
operation
returns
enters
the MIN
SETUP
function.
This sets up
the front-panel
controls
as the Auto-Setup
does
but leaves
many
of
the
controls
unchanged
from
what
the user
has
previously
set.
MIN
SETUP
action for each
front-panel control Is shown
in Appendix
A.
2247A Operators
2-25